v1rev 52017-12-07
total performance (diff. press., 69 bar static pr., 28°C, overpr., zero & span)
0112/2///61987#ABB943
for differential pressure measuring devices, square root of: nonlinearity^2 + (maximum ambient temperature error per 28 °C change)^2 + (maximum static zero point error per 69 bar change)^2+(maximum static span error per 69 bar change)^2 +(the static overpressure effect on one side of a differential pressure)^2, expressed as percentage of span
- Used by classes
- 1
- Code
ABB943- Data type
- REAL_MEASURE_TYPEmeasure
- Unit
- UAA000 UAA000
All properties31#
- Version
- 1
- Revision
- 5
- Preferred name
- total performance (diff. press., 69 bar static pr., 28°C, overpr., zero & span)
- Definition
- for differential pressure measuring devices, square root of: nonlinearity^2 + (maximum ambient temperature error per 28 °C change)^2 + (maximum static zero point error per 69 bar change)^2+(maximum static span error per 69 bar change)^2 +(the static overpressure effect on one side of a differential pressure)^2, expressed as percentage of span
- Note
- Definition class
- 0112/2///61987#ABA000Equipment for industrial-process automation
- Unit structure
- %
- Initial value
- 0112/2///62720#UAA000 UAA000
- Responsible committee
- IEC/SC 65E
- Publisher
- IEC
- Published in
- IEC 61987
- Published in
- NON_DEPENDENT_P_DET
- Status level
- Standard
- Version release date
- 2026-05-04
- Revision release date
- 2017-12-07
- Version initiation date
- 2026-05-04
- Original definition date
- 2017-12-07
- Change request ID
- C00200
- C017
- C018
- C019
- MDC_P001_6
- 0112/2///61987#ABB943total performance (diff. press., 69 bar static pr., 28°C, overpr., zero & span)
- MDC_P022
- REAL_MEASURE_TYPE